Intrinsically Safe for Class I, Division 1, Groups ABCD; Non-Incendive for Class I, Division 2, Groups ABCD Locations
316 Stainless Steel Enclosure
Compact Design
Fast Response Time
Faster Downloads (1 Minute Full Memory)
Docking Station Communication
Easily Identify Low Battery Voltage
Ready to Use Right Out-of-the-Box
Free NIST Certification
Description
This product has been replaced by the OM-CP-RHTemp1000IS-A2 which has the same features, price, and is certified to the same Intrinsically Safe standards.
The new OM-CP-RHTEMP1000IS-A is an intrinsically
safe temperature and humidity data logger certified for
Class I, Division 1, Groups A, B, C, and D and nonincendive
for Class I, Division 2 Groups A, B, C, D
environments. This certification makes the device ideal
for EtO sterilization, environmental studies, medical and
pharmaceutical and numerous other hostile environment
applications.
One of the enhancements to the OM-CP-RHTEMP1000IS-A
is the stainless steel enclosure with engraved serial
number and device name. The device interfaces with
the OM-CP-IFC400 docking station or the
OM-CP-IFC406 multiplexer data logger interface
(sold separately) and can be placed directly into
the dock for full communications. No more removal
of end-caps to download data. The small size of the
OM-CP-RHTEMP1000IS-A allows it to be placed in
remote or hard to reach locations. The device now
features a battery life indicator allowing users to easily
identify low voltage and change batteries when required.
This feature maximizes the use of your battery life and is
a cost saving efficiency.
The OM-CP-RHTEMP1000IS-A utilizes the latest
software. This allows for simple starting, stopping and
downloading of collected data. The devices are able
to store up to 16,350 time and date stamped humidity
and temperature readings in non-volatile solid state
memory. Once the readings have been downloaded to
the software, it can be viewed in graphic, tabular, and
summary form for easy analysis, as well as the potential
to be exported into Excel® for further calculations.
Specifications Temperature Temperature Sensor: Resistance Temperature Detector (RTD) Temperature Range: -40 to 80°C (-40 to 176°F) Temperature Resolution: 0.01°C Temperature Response Time in Air:
t63 - 8 minutes to 63% of change
t90 - 68 minutes to 90% of change Calibrated Accuracy: ±0.5°C (0 to 50°C) Humidity Humidity Sensor: Capacitive polymer Humidity Range: 0 to 100% RH (non-condensing) Humidity Resolution: 0.1% RH Humidity Response Time in Air:
t63 - 30 seconds to 63% of change
t90 - 2 minutes to 90% of change Calibrated Accuracy: ±3.0% RH maximum
General Start Modes: Software programmable immediate start or delay
start up to 24 months in advance Real Time Recording: May be used with PC to monitor and
record data in real time Memory: 16,350 readings per channel Password Protection: Yes Reading Rate: 1 reading every second up to 1 reading every
24 hours Calibration: Digital calibration through software Calibration Date: Automatically recorded within device
Memory Wrap: Yes Data Format: Date and time stamped °C, °F, K, °R ; %RH,
mg/mL, dew point Battery Life: 2 years typical at 25°C (15 minute reading rate) Battery Type: 3.6V lithium battery included, user replaceable Time Accuracy: ±1 minute/month at 25°C Computer Interface: OM-CP-IFC400 USB docking station or
OM-CP-IFC406 multiplexer interface required; 125,000 baud Software: Windows® XP SP3/Vista/7 and 8 (32-bit and 64-bit) Operating Environment: -40 to 80°C (-40 to 176°F),
0 to 95% RH (non-condensing) Dimensions: 42 L x 24.6 mm dia (1.7 x 0.97") Weight: 65 g (2.3 oz) Material: 316 Stainless Steel
† All amounts shown in AUD Note: Comes complete with 3.6V lithium battery. Operator’s manual and USB interface cable are included with the OM-CP-IFC400 software/cable
package or OM-CP-IFC406 multiplexer (required for data logger operation), sold separately.
This product page contains items that either have a Lithium Battery installed in the equipment or have a Lithium Battery packed with the equipment. These items as being regulated for air transportation. The following applies:
• No FX1D, FX2D or FX3D can be used as a ship method.
• Only UPS1, UPS2 or UPS3 is authorized.
• For International shipments UPS and DHL can be used. For DHL internationally – Omega only has permission to ship Lithium Batteries no other Hazardous Material.